Step Key Display
1
Open the parameter
menu.
2
Step through the menu
to the STD 1 parameter.
3
Move across to the set-
ting fi eld.
4
Decrement to required
level.
5
Return to parameter
fi e l d .
6
Step to the STD 2 pa-
rameter.
7
Move across to the set-
ting fi eld.
Entering reference block values
Step Key Display
8
Increment to required
level.
9
Return to normal mea-
surement mode.
Normal measure-
ment display
User reference blocks
Can be used to set the AutoSigma 3000 DL, provided
that
• STD 1 is within the range 30 to 63 % IACS,
• STD 2 is within the range 6 to 50 % IACS and
• STD 1 is at least 10 % IACS above STD 2.
The values are entered in the unit in the same way as
the STD blocks.
Considerations
If user reference blocks are used, accuracy can be
compromised and range limited.
Operation
AutoSigma 3000 Issue 04, 06/2007 5 - 5