6.1 Connection
The 7 way Lemo Data Output socket is provided for
connection to Data Handling Devices, Printers and
PCs.
The communication link protocol depends on the cable
connected using an ASCII Data transfer. A Mitutoyo
data cable is automatically sensed so an external Data
Logger can request a reading.
If printers or PCs etc that use serial RS-232 format are
to be connected, converter leads are required. Product
codes: 47A004 (PC), 47A002 (Printer)
Signal identifi cation of connections to the 7 Pin LEMO
Data Output socket.
Pin 1 0 V
Pin 2 +5 VA
Pin 3 TXDATA1
Pin 4 ~CTSREQ
Pin 5 RXCLOCK
Pin 6 0 V
Pin 7 ~MITSEL
Connection
External data handling devices
6 - 2
Issue 04, 06/2007 AutoSigma 3000