■ Single Lead vs. Multi-Lead:
This option examines ECG
leads utilized for arrhythmia
analysis.
• Multi-Lead Analysis examines
ECG leads I, II, III and V to
help eliminate false alarms
and improve the ability of the
system to:
– Detect beats that occur
isoelectric to a single chest
lead.
– Discriminate artifact that
appears in one lead
compared to other lead
vectors.
– Provides a “Smart Lead
Fail” feature where the
failed lead is identified and,
if available, another lead is
provided for display.
– Continue arrhythmia
processing after a lead
change.
• Single Lead Analysis uses
only the top displayed lead
to process heart rate and
arrhythmia information.
• Single lead analysis can be
beneficial when
troubleshooting heart rate
and arrhythmia alarms.
• To Change to Single Lead
Analysis:
– Select the ECG Parameter
Window.
– Select LD ANALYSIS:
MULTI-LEAD.
– Move the cursor in front of
Single Lead and press to
select.
– Select MAIN MENU to
exit.