12 Lead ECG
12 Lead ECG Analysis
NOTE: For the most accurate serial comparison, use the same
electrode configuration used on previous analysis.
■ To Obtain a 12-Lead on a
Patient:
• Place patient in a supine
position.
• Correctly identify and apply
all 10 electrodes.
• Select the ECG Parameter
Window.
• Select 12 LEAD ECG
ANALYSIS.
NOTE: Accurate demographics must be entered to ensure
proper analysis.
• Select ADMIT INFO
• Select CHANGE ADMIT
INFO.
• Enter Information:
– First Name: Completely
entered.
– Last Name: Completely
entered.
– Patient ID: Facility Specific,
(i.e., Medical Record
Number, Social Security
Number, etc.)
– Sex
Electrode Placement
– Birth Date
– Age: Correct age impacts
the analysis.
– Height
– Weight
– Location ID: Identifies
origin of 12-lead to MUSE
for transmission.
(Set in defaults. For rover
modes, set in admit menu).
– Site Number: Identifies
hospital to MUSE for
transmission.
(Set in defaults).
NOTE: Depending on facility policy, other fields may be
required; i.e., Technician ID, Optional Field, Order Number.
(continued on next card)
V1 Fourth intercostal space at the right
sternal border; right chest.
V2 Fourth intercostal space at the left
sternal border; left chest.
V3 Midway, between locations V2 and
V4; left chest.
V4 Mid-clavicular line in the fifth
intercostal space; left chest.
V5 Anterior axillary line on the same
horizontal level as V4; left chest.
V6 Mid-axillary line on the same
horizontal level as V4 & V5; left chest.
LA Left deltoid or left wrist.
RA Right deltoid or right wrist.
LL Left thigh or left ankle.
LA Right thigh or right ankle
Front
Leads
12 Lead Electrode Placement