Masimo SET allows the
sensitivity and averaging times
to be adjusted:
Sensitivity options include:
Normal, for routine patient
monitoring purposes.
Maximum for improved low
perfusion performance and
for faster tracking of SPO
2
changes.
Averaging menu options
determine how many
seconds the collected SPO
2
information is averaged.
Options for averaging times
are: 2, 4, 8, 10, 12, 14, or
16 seconds.
Rejuvenating the Adhesive
Properties of the Disposable
Saturation Probe:
Rejuvenate the probe at least
once per shift and PRN.
Rub the adhesive sides of the
probes with an alcohol
preparation pad.
After the probe has dried for
at least one minute, replace it
on the site.
NOTE: It is recommended that the adhesive tapes be changed
with every site change. Adhesive tape can not be rejuvenated.