■ Alarm Volume: The alarm tone
volume can be adjusted at the
bedside.
• To Adjust Alarm Tone
Volume:
– Select MORE MENUS.
– Select ALARM
CONTROL.
– Select ALARM VOLUME.
– Turn the Trim Knob to the
desired alarm tone and
press to select.
– Select MAIN MENU to
exit.
NOTE: The alarm volume at the bedside monitor does not
affect the alarm volume at the central station.
NOTE: A minimum alarm volume can be set up in the monitor
defaults.
■ Clear Alarms: Allows for any
alarm information displayed in
the alarm parameter window to
be cleared from the display.
• To Clear Alarms:
– Select More Menus.
– Select Alarm Control.
– Select Clear Alarms.
– Information from the alarm
parameter window is
removed.
NOTE: Arrhythmia alarms are not deleted and can be found in
the Alarm History.
■ Alarm History: See Patient
Data Tab.
■ Display Off/Alarm Pause: This
option allows the user to
disconnect the patient from the
monitor for an extended period
of time. There are several alarm
pause choices available:
• Display Off/Alarm Pause:
This option turns off the
bedside monitor and pauses
the alarms at both the bedside
and the central station for an
indefinite period of time.
• Monitor Pause: This option
turns off the bedside monitor,
but allows alarm notification
to remain active at the central
station.
• To Activate the Display Off/
Alarm Pause Feature:
– Select MORE MENUS.
– Select ALARM
CONTROL.
– Select DISPLAY OFF/
ALARM PAUSE.
– Move the cursor in front of
the desired alarm pause
choice and press to select.
– The display screen is now
paused.
– Press the Power button on
the monitor to reactivate
the display and alarms.
NOTE: Pressing the Power Button will turn the Dash display off
and stop communication with the Central Station. Depending
on your configuration, a NO COMM alarm may sound.