Page #57 background imageLoading...
Page #57 background image
TAB
TRIM
COVER &
NON-TAB
TRIM
TAB
TRIM
COVER &
NON-TAB
TRIM
All Limits: This information
window shows all the high and
low alarm limits for all
parameters that are currently
being monitored.
To Set Parameter Alarm
Limits:
Select MORE MENUS.
Select ALARM
CONTROL.
Select ALL LIMITS.
An Information Window is
displayed.
Turn Trim Knob to move
the cursor in front of the
desired parameter that
requires an alarm limit
adjustment and press to
highlight parameter label.
Turn Trim Knob to
highlight HIGH LIMIT or
LOW LIMIT and press to
select.
Turn Trim Knob to change
limit and press to select.
Turn Trim Knob to
highlight PARAMETER
LABEL and press to select.
Move the cursor in front of
RETURN to close window
and press to select and
close window.
Select MORE MENUS.
Select MAIN MENU to
exit.
NOTE: Parameter alarm limits can also be adjusted in the
individual parameter menus.
NOTE: Any changes made with the Parameter Alarm Level
menu options are temporary and return to the default settings
upon patient discharge.