Respirations /
Temp
■ Sensitivity: The monitor
automatically sets the
detection sensitivity at 40% of
the average amplitude. The
sensitivity may need to be
changed due to varying
amplitudes or artifact. The
lower the percentage, the
greater the detection
sensitivity.
• To Change the Sensitivity:
– Select RESPIRATIONS.
– Select SENSITIVITY.
– Turn the Trim Knob to the
desired percent and press
to select.
– Select MAIN MENU to
exit.
NOTE: The RELEARN option will return the sensitivity to 40%.
■ Respiration Limits:
• To Change the Respiration
Limits:
– Select RESPIRATIONS.
– Select the desired
respiratory rate limit.
– Turn the Trim Knob to the
desired respiratory rate and
press to select.
– Select MAIN MENU to
exit.