Page #52 background imageLoading...
Page #52 background image
TAB
TRIM
COVER &
NON-TAB
TRIM
TAB
TRIM
COVER &
NON-TAB
TRIM
Cardiac
Output
When the INJECT WHEN
READY is displayed, proceed
with injection.
Seconds after injecting,
COMPUTING CO is
displayed.
CO COMPLETE is displayed
with the CO value upon
completion and the data is
placed in the trial window.
NOTE: Up to four trials are displayed in the trial window, the
fifth trial automatically deletes the first. All trials are
averaged and the data is stored in the cardiac calcs.
Select DELETE CO TRIALS.
Move the cursor in front of
the trial that is to be deleted
and press to highlight.
Move the cursor to RETURN
and press to select.
NOTE: Trials are deleted permanently.
Select CARDIAC CALCS.
Select CHANGE VALUE to
edit or add unmonitored
parameter values.
Move the cursor to RETURN
and press to select.
Select SAVE CALCS.
Select REVIEW CALCS.
Select MAIN MENU to exit.
NOTE: The last average CO value, along with a time stamp,
will be displayed in the parameter window.