EasyManua.ls Logo

GE DASH 3000 V5

GE DASH 3000 V5
105 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TAB
TRIM
COVER &
NON-TAB
TRIM
TAB
TRIM
COVER &
NON-TAB
TRIM
SPO
2
Problem:
Why does the monitor not display an SPO
2
reading after
changing the Masimo probe site?
Answer:
The Masimo probe is not recognized by the monitor.
Intervention:
When a Masimo sensor is repositioned at any time,
disconnect the cable from the sensor before
repositioning.
Reconnect the sensor to the interface cable after
proper patient preparation and placement.
Problem:
Why is the SPO
2
parameter window not displayed after
connecting the SPO
2
interface cable and probe?
Answer:
No SPO
2
data is displayed due to a hardware failure, or
an unrecognized or defective probe.
Intervention:
Ensure the probe is attached to the interface cable
and the monitor.
Change the probe.
Change the cable.
If the issue remains, contact the hospital Biomedical
Engineering Department for service.
TROUBLESHOOTING and
FREQUENTLY ASKED QUESTIONS

Table of Contents

Related product manuals