EasyManua.ls Logo

GE DigitalFlow DF868 - Page 57

GE DigitalFlow DF868
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programming Site Data 1-49
March 2008
Setting Up Alarms Complete the following steps to set up the three alarm relays of an
option card installed in Slot x:
1. Press
[F1]-[F3] to set up alarm relays A, B, or C, respectively.
Note: The set up of alarm A is used as an example. Identical
procedures would be used to set up the other alarms.
2. Press
[F1]=OFF to disable Alarm A and return to the previous
prompt, or press
[F2]=HIGH, [F3]=LOW, [F4]=FAULT to specify Alarm
A as the type indicated.
3. Press
[F1] for standard operation or [F2] for Failsafe operation. See
Chapter 1, Installation, of the Startup Guide for wiring
instructions for each mode.
4. Use the [←] and [→] and
[F1]-[F4] keys to select a Channel option
for the data source, as shown in Table 1-20 on page 1-47.
5. The program now varies, depending on your selection in step 2.
If you selected FAULT as the alarm type, press [F1]-[F3] to select
the type of error (flow, non-flow, or both) you wish to trigger the
alarm.
Note: If you have enabled the Energy Option, the DF868 treats the
E8 and E9 errors (supply and return temperature) as flow
errors.
If you selected HIGH or LOW:
a. Use the [←] and [→] and
[F1]-[F4] keys to specify the desired
Output Measurement parameter. See Table 1-21 on page 1-47
for a description of the available options.
Note: The measurement units that appear in these prompts are those
selected in the GLOBL-SYSTM menu earlier in this chapter.
b. Enter a value for the trigger point of the alarm and press
[ENT].
c. Press
[F1]-[F3] to set up another alarm or press [EXIT] to select
another slot for set up.

Table of Contents

Other manuals for GE DigitalFlow DF868

Related product manuals