Girae Warmer 1-16
Screen
Step
Description
3 Patient Temperature Alarm
Push the “Pt. Temp Alarm” key to choose
between two settings in the warmer’s
Baby Mode. The 0.5 setting activates
an alarm when the dierence between
the set Temperature and the baby’s skin
temperature is greater than 0.5°C. The
default 1.0 setting activates an alarm when
the dierence between the set temperature
and the baby’s skin temperature is greater
than 1.0°C.
4 Manual Temperature Alarm
You can use the manual temperature alarm
when you wish to manage the baby’s
temperature in the Manual Mode, but want
to be notied when the baby’s temperature
drops below or rises above a threshold
setting.
Place the temperature probe on the baby’s
skin. Push the “Man Temp Alarm” key to
choose from 3 settings in the warmer’s
Manual Mode. The default O setting
deactivates the alarm, the 0.5 setting
activates an alarm when the dierence
between the set temperature and the baby’s
skin temperature is greater than 0.5°C. The
1.0 setting activates an alarm when the
dierence between the set temperature and
the baby’s skin temperature is greater than
1.0°C. For more details on operation of the
Manual Temperature Alarm, (see “Setting the
manual temperature alarm” in chapter3).
NOTE: After power down, only the time setting will remain.