3-25Girae Warmer
Nurse Call checkout
Steps Description
1
Complete the basic operating procedure in chapter 3.
2
Verify proper operation of the Nurse Call station.
3
Connect the Nurse Call connector to the warmer.
4
Place the unit into Baby Mode and unplug the patient probe to trigger an alarm. Verify that
you also get an alarm at the Nurse Call station.