1-21Girae Warmer
Check Baby-Heat O
Screen
Step
Description
1
Baby Mode
A continuous tone sounds after the radiant
heater has been running at 100% power for
more than fteen minutes, and the
“Check Baby” alarm has gone unanswered
for three minutes. The radiant heater then
shuts o automatically and the high priority
“Check Baby-Heat O” alarm activates.
Manual Mode
A continuous tone sounds after the warmer
has been operating at greater than 25%
heater power* for more than fteen minutes,
and the “Check Baby” Alarm has gone
unanswered for three minutes. The radiant
heater then shuts o automatically and the
high priority “Check Baby-Heat O” alarm
activates.
Action: silence the alarm and check the
attachment of the skin probe to the baby
and monitor the baby’s temperature closely.
*Default power setting is 25%, but this can be adjusted
on the Service Screen. See service manual.
Conrm Probe Jack Connection
Screen
Step
Description
1
Baby Mode
The “Conrm Probe Jack Connection” alarm
activates when the skin probe jack is un-
plugged.
Manual Mode
It will also activate when there is no probe
in the jack in the Manual Mode, if the Man
Temp Alarm has been set.
Action: Check that the skin probe is fully
inserted into the skin probe jack If this does
not cancel the alarm, check if the Man Temp
Alarm has been set. If the alarm is still not
cancelled, then replace the temp probe.