4-5Girae Warmer
Cleaning Solutions
Cleaning solutions that may be used safely:
Generic Formulation Maximum Concentration
Sodium Hypochlorite (bleach) 0.5% Aqueous Solution
Glutaraldehyde 2%
Hydrogen Peroxide 6%
Iodophor Solution 0.27%
Cavicide
®
100% spray
Do not use the following cleaners; they will damage the parts you are cleaning and are not
recommended:
• Isopropyl Alcohol (in concentrations greater than 15%)
• Quaternary Ammonium (such as Virex)
• Solvents (such as acetone)
CAUTION:
Use of cleaning/disinfecting solutions containing chemicals not listed above, i.e. alcohol,
acetone, etc., or chemicals in greater concentrations than those listed above, may damage
the probe.