Girae Warmer 2-6
Illustration
Step
Description
6
Unplug the skin probe and change the
warmer to Baby Mode. Check that both
visual and audio alarms trigger in the Baby
Mode. Silence the alarm by using either the
alarm silence key or the Hands Free Alarm
silence.
7
Return the warmer to Manual Mode by
pressing the “Mode” key. Select a Heater
Power setting to silence the prompt tone.
8
If the unit is equipped with an elevating
bed, check the operation of the bed eleva-
tion mechanism. Raise and lower the bed
along its entire travel range, checking that
the mechanism operates smoothly. Check
that the two bed height adjustment pedals
on either side of the unit raise and lower the
bed height.
NOTE: If readying the bed to admit a new patient, power down the warmer for 10 seconds,
then turn power back on. Allow warmer to preheat the bed in Warmup Mode.