9.11.11 Codonics Horizon Print Parameters
Table 12-25 Codonics Horizon Print Parameters
TAB Attribute Value Comments
Printer DICOM Printer Label HORIZON Or customers preference
Application Entry Title Enter value given by Site Network
Administrators: PRINT_SCP
AE Titles may be site
specific, be sure to check
on-site
IP Address Enter value given by Site Network
Administrator
Port Number Enter value given by Site Network
Administrator, eg: 104
Port numbers may be site
specific, be sure to check
on-site
Pixel Depth 12
Printer Pixel Size (micron) 79.5
Configuration Information LUT = LINEAR
Density Min. = 0, Max. = 300
Magnification Type Replicate
Smooth Factor 0 Not supported
Trim NO
Polarity NORMAL
Border Density BLACK Can also pass OD value of
(0 - 310)
Empty Image Density BLACK Can also pass OD value of
(0 - 310)
Printer Memory Size 40Mbytes
Layout Select the desired layouts
allowed for this printer
Slide formats

Table of Contents