Chapter 10 - - Renewal parts Page 245
GE H
EALTHCARE
D
IRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Section 2.0
Digital Detector
The device or portions of the device may have power present from the batteries even when shut
down, please demount the battery first.
Detector Assembly
Before the replacement, make sure detector’s malfunction is not caused by collision.
1.) Open the service port at the back of detector.
2.) Check the vibration sensor.
• If the center of the vibration sensor is Transparent, it means no any collision during usage.
• If the center of the vibration sensor is RED, it means collision exists during usage. Please
inform the user to carry the detector carefully.
3.) After the detector assembly replaced, please refer to Detector Association on page 117 to
register and associate the new detector with system.