GE HEALTHCARE
DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Chapter 4 - - Target System Check Page 43
Chapter 4 - Target System Check
Digital X-Ray Imaging System (Brivo XR118) can upgrade target analog system which meets the
below requirements to DR system.
Before installation XR118, purchaser needs to check target system from two sides.
1.) In order to reduce installation time, please call customer to check target system before go to
site, including:
a.) The system can be installed into Wall stand/Table. Make sure the cassette on target
system can hold the detector (Width: 383.5±1mm, Height:459.5±1mm).
b.) Target system can make exposure.
c.) Target system is fix RAD system.
d.) Target system’s Generator frequency is higher than 400Hz.
e.) Suggest the hospital to ask RIS engineer also to be on site for DICOM connectivity in case
RIS query issue occurs, it often needs RIS side’s update to meet DICOM3.0.
2.) Make sure the target system meets all the requirements below. For more information, please
contact your sales representative.
Table 4-1 Target System Requirements
No. Categor
y
Item Preferred Capability Target
System
Claim
Pass
/Fail
1 Generat
or &
Tube
Maximum kV >=80kV, kV accuracy <=10%
2 Maximum mA >=50mA, mA accuracy <=20%
3 Maximum mAs >3.2mAs, mAs accuracy <= ±
(10% + 0.2mAs)
4 Maximum exposure
time
<=10s, exposure time accuracy
<=± (10% + 1 ms)
5 Generator inverter
switching frequency
>=400Hz
6 Generator Comply with GB9706.3-2000/
IEC60601-2-7:1998 or IEC
60601-2-54: 2009
7 Tube Comply with GB9706.11-1997/
IEC601-2-28:1993 or IEC
60601-2-28: 2010
8Cassette
Tray
Size Compatible with ISO4090
35cmx43cm, Thickness 16mm
cassette
9 Grid Line Pair 36, 40, 60, 70, 78lp/cm