DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Page 214 Section 4.0 - Manual Bad Pixel Calibration
Section 4.0
Manual Bad Pixel Calibration
1.) Insert the service key, launch the SUIF.
2.) Quit SUIF
3.) Open the raw image in image viewer to mark the bad line or bad pixel
Note: To mark bad line, use the line mark tool to completely overlap the bad line.
To mark the bad pixel, use the ellipse mark tool to just cover the bad pixel area.
4.) Click BP button.
5.) Click ok to continue.