GE HEALTHCARE
DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Page 48 Section 2.0 - Cables
Section 2.0
Cables
Note: All system cable connection, please refer to the marks on the cable.
2.1 Routing
• System MIS cables should be pulled to/from the System Components before the System
Components is moved into the final position.
• Some system cables are pre-connected to a sub-system and should not be disconnected in
order to pull the cable(s) through the conduits or wall ducts.
• Power cables must be kept separate from signal cables as much as possible. In wall ducts or
raceway, power cables should be routed in a separate channel from signal cables. Power
cables should not be bundled together with signal cables.
All MIS cables should be routed through the conduits or wall ducts as rat-proofed as possible.
2.2 System Cables
1.) Perform routing and connection of system and power cables.
2.) Refer to the schematics for details on system cabling.
System cables may cause a trip hazard.