GE HEALTHCARE
DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Chapter 6 - - Calibration Page 143
Section 6.0
Detector QAP
6.1 Before You Begin
The following applies when performing Quality Assurance Procedure (QAP).
Follow the instructions presented on SUIF.
6.2 When to Perform QAP
On a scheduled, weekly basis or QAP required icon is shown. And perform QAP every time
you calibrate the detector and as a final system check following system installation. For detailed
Detector QAP procedure, please refer to the QAP chapter of Operator Manual.