GE HEALTHCARE
DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Chapter 6 - - Calibration Page 143
Section 6.0
Detector QAP
6.1 Before You Begin
The following applies when performing Quality Assurance Procedure (QAP).
• Follow the instructions presented on SUIF.
6.2 When to Perform QAP
On a scheduled, weekly basis or QAP required icon is shown. And perform QAP every time
you calibrate the detector and as a final system check following system installation. For detailed
Detector QAP procedure, please refer to the QAP chapter of Operator Manual.