DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Page 74 Section 9.0 - System Configuration
Section 9.0
System Configuration
9.1 When to Perform System Configuration
Potential for Data Loss:
1.) Always create an up-to-date backup of your system data following product installation or
changes to configuration data.
2.) Upon exiting the Service User Interface (SUIF), for Configuration and Calibration information
to take effect, the system must be reset. This must be done before the system is used for
patient acquisitions.
9.1.1 During System Installation
Your system is delivered to you configured to default values. You should configure your system to
the values that are appropriate for your particular installation. The items that must be configured are
identified in Table 12-1.
Table 12-1 System Configuration Parameters
9.1.2 Missing or Lost System Backups
If you perform a software install and do not restore configuration files, you must re-perform all of
configurations listed in, after it’s complete. A software install resets all configuration data. If you
performed a [backup] before you began the software install, you are not required to perform the
configuration and calibrations. Simply [restore] the data saved previously.
9.1.3 Addition of Purchased Options
Installation of purchased options can only be performed by your authorized GE sales and service
representative. If you add purchased options to your system, you should update the system
configuration files immediately and create and save a new backup.
Comments System Configuration Parameter
See System Configuration Tool
Parameters
Site
General
See Hardware AP Configuration Wireless Connection
See DICOM Configuration & DICOM
Printers
DICOM
See Operator Manual (OM) Installation Operator Manual
See Adding a Bar Code Reader Bar Code Reader (Option)
See User Accounts System Login (HIPPA)