GE HEALTHCARE
DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Chapter 5 - - Installation & Configuration Page 83
9.8 Operator Manual (OM) Installation
The operator manual must be installed:
- when the system is initially installed
- following a complete system load from cold
- following a software reload of the workstation PC
The procedure follows:
1.) Verify that no archival process is running.
2.) Verify that system is not in a state where Service Browser is invoked.
3.) Press the following keys: <Cntrl><Alt><F2> simultaneously, to open a command window.
4.) At the login prompt, type: xruser
5.) Enter the password: 4$xray
6.) Insert Operators manual CD into the tray of DVD drive PC
7.) To run the inter-active script, at prompt type the following: ./xrdinstall.om, click ENTER key.
8.) A message is displayed: Do you want to proceed? Enter "y", click ENTER key.
9.) Below message is displayed:
Select the language to install:
1 for English
2 for French
3 for Spanish
4 for Chinese
5 for Turkish
6 for Indonesian
7 for Russian
8 for Portuguese
9 for Korean
10 for All Languages
11 for Quit, no installation shall happen
Enter "10" for Enter Choice. Click ENTER key.
10.) A message is displayed: Installing Online Manual in all languages.
Installation Complete will be displayed after OM installation.