GE HEALTHCARE
D
IRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Page 228 Section 2.0 - Preventive Maintenance Action List
Section 2.0
Preventive Maintenance Action List
Operation Frequency
(Month)
Duration
(Min)
Description
1.) General
Review Previous
Performance
12 15
A.) Customer Discussion - Discuss the objectives of the
PM with the customer: The purpose of planned
maintenance is inspection and not repair.
i.) Discuss room scheduling the PM.
ii.) Special considerations for patient needs.
iii.) Housekeeping concerns and needs.
B.) Logbook Review - The site log is a history of the
system and room’s performance. The system and
room should be in good functional condition. If there
are any major problems or concerns, it’s suggested
that PM be deferred to a later time and date.
C.) Error Log Review - The error log is a history of the
system’s performance. Review the log for any errors,
trends or unusual events.
D.) Cable Connection Check - Check all cables are not
damaged and securely connected.
Check Protective Circuits 12
15
Check interlocks (room and door interlocks), e-stops and
exposure-on warning lights on control console.
Exposure check 12 Take exposure. Verify system is working well.
2.) Detector
Detector power on button
function check
12
15
Ensure button is in good condition and making good.
Detector appearance check 12 Clean and inspect.
Detector battery check 12 Check battery appearance and clean.
3.) AP
AP appearance check 12 5 Clean and inspect.
4.) Dock
Dock appearance check 12 5 Clean and inspect.
Dock connection check 12 5
Insert USB mouse in every USB port to check connection
status.
5.) Workstation