Chapter 9 - - Periodic Maintenance Page 229
GE H
EALTHCARE
D
IRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Workstation 12
30
A.) Clean and inspect.
B.) Fix hard disk to keep connection tight into
workstation. Refer to Portable workstation SSD drive
replacement procedure on page 238.
C.) Clean up disk space.
D.) Check workstation battery appearance and capacity.
Keyboard 12 Clean and inspect
Monitor 12 Clean and inspect
Mouse 12 Clean and inspect
USB DVD driver 12
Perform system backup operation to check USB DVD
driver performance.
Barcode Reader 12 Clean and inspect
6.) Calibration
Detector BP, Gain
calibration
12 20
Perform detector Gain, Bad Pixel. Follow instructions in
SUIF.
7.) System Performance
Evaluate Image Quality 12 10
Perform the Quality Assurance Process (QAP). Follow the
instructions in SUIF.
Operation Frequency
(Month)
Duration
(Min)
Description