GE HEALTHCARE
DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Page 62 Section 8.0 - Software Installation
Section 8.0
Software Installation
8.1 Before You Begin
Potential for Data Loss. Archive all protocols, patient files, and backup calibration and configuration
data. It's recommended that a system backup be performed prior to any installation, upgrade or
restoration of software. The installation of software can destroy existing files, including patient,
calibration and configuration files. You might have to configure and calibrate your system, if you do
not have a good backup to restore!
Note: Configure the Time Zone after software re-installation or upgrade.
Note: The system is delivered with software installed. Only software configuration is required for the new
site installation.
8.2 System Data Backup
Files within the workstation need to be saved, if system software is reload, upgraded or becomes
corrupted. After system installation, saved files can be restored in their original condition and
location.
8.2.1 System Backup’s Coverage
The following data is part of a system backup:
All Calibration Files
All Configuration Files.
InSite Checkout Downloaded Files
Custom Looks
HIPPA Login and user accounts
DICOM Printers/Preferences/Network Configurations.
Detector Management files: Could be backup only after QAP or detector calibration is finished.