DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Page 86 Section 9.0 - System Configuration
9.10 DICOM Printers
Note: For DICOM printer, please acquire mandatory parameters (Application Entry Title \ Network
Address \ Port Number) from hospital supervisor.
For other parameters, please configure exactly according to the DICOM printer data sheet.
After configured, click “C-Echo Test” to confirm the DICOM printer connectivity.
Table 12-9 Printer Tab
Table 12-10 Layout Tab
Table 12-11 Film Size Tab
Table 12-12 Printer Settings Tab
Attribute Host Device 1 Host Device 2 Host Device 3