EasyManua.ls Logo

GE Healthcare Brivo XR118 - Page 119

GE Healthcare Brivo XR118
318 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Chapter 5 - - Installation & Configuration Page 119
4.) Enter the new Detector Name.
5.) Chose to make the new Detector Associated or Non-Associated.
6.) Press [Save]. The new Digital Detector is now registered with the system.
7.) Label the new Digital Detector with the Detector Name.
10.2.2 Adding A New Detector Identified By Color and Shape
Figure 5-17 Add a detector By Color and Shape
1.) Press the [Add] button on the Detector Registration screen.
2.) Enter new detector’s MAC address.
Note: Please detach the detector battery to find the mac address.
3.) Chose Color and Shape for the Select Identification Type.
4.) Chose the Color and Shape from the drop-down list.
5.) Chose to make the new Digital Detector Associated or Non-Associated.
6.) Press [Save]. The new Digital Detector is now registered with the system.
7.) Attach the plastic tag with the appropriate shape and color onto the Digital Detector.

Table of Contents