GE HEALTHCARE
DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Chapter 5 - - Installation & Configuration Page 119
4.) Enter the new Detector Name.
5.) Chose to make the new Detector Associated or Non-Associated.
6.) Press [Save]. The new Digital Detector is now registered with the system.
7.) Label the new Digital Detector with the Detector Name.
10.2.2 Adding A New Detector Identified By Color and Shape
Figure 5-17 Add a detector By Color and Shape
1.) Press the [Add] button on the Detector Registration screen.
2.) Enter new detector’s MAC address.
Note: Please detach the detector battery to find the mac address.
3.) Chose Color and Shape for the Select Identification Type.
4.) Chose the Color and Shape from the drop-down list.
5.) Chose to make the new Digital Detector Associated or Non-Associated.
6.) Press [Save]. The new Digital Detector is now registered with the system.
7.) Attach the plastic tag with the appropriate shape and color onto the Digital Detector.