EasyManua.ls Logo

GE Healthcare Brivo XR118 - Page 167

GE Healthcare Brivo XR118
318 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Chapter 8 - - Troubleshooting Page 167
1.3 Wireless Debug
1.3.1 Detector connection lost
1.) Go to SUIF -> Diagnostics -> Wireless Link.
2.) Click “Start” button.
3.) The wireless link diagnostics result and the recommended debug actions will be displayed.
4.) Follow the instructions on the screen to debug the wireless issue.
5.) Other recommended instructions:
- Ping AP IP address 3.1.1.1 to make sure the AP is connected, then reboot the detector.
- Or release the detector then associate the detector again.