GE HEALTHCARE
DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Page 182 Section 2.0 - Software Issues
The steps to handle this special case are following.
- Login with the root’s password #superxr
- Input “fsck –y” to fix the crashed file system automatically.
- After above operation done, input “reboot” to restart the Workstation.
- Take a snapshot for further investigation on the root cause.
10.) How to take screens hot?
i.) Input iserver external IP and connect the iServer by putty software.
ii.) In the opened X command page, enter user and password
iii.) Enter command " xterm -display :0 -geometry 120x50 -e /magichome/xruser/engg/
engg_ui & "