DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Page 218 Section 4.0 - Manual Bad Pixel Calibration
12.) Click "Continue" button.
13.) Click "Overwrite", then click "Continue" button.
Note: Either overwrite or append is ok for first time of MBP, suggest choose Append for second time
MBP.