1.2 LOTO Procedure for System Power
Name of Equipment: System
Step Lockout/Tagout Procedure
1 Prepare for shutdown of equipment by notifying affected personnel working in the area
that lockout/tagout is being performed.
2 Bring system software down. See De-Energizing System on page 26. And return to step
3.
3 Locate main power supply (Service) disconnect feeding System Cabinet.
4 Lock Out / Tag Out the main power disconnect.
5 Using an insulated voltmeter, measure the incoming power connection to the system
cabinet at the circuit breaker.
6 If no voltage is present at the system cabinet’s circuit breaker; place the power switch on
the Cabinet to the OFF position and service equipment as required.
7 After service has been performed, notify affected personnel that equipment will be re-
energized and LOTO devices are being removed.
8 Assure the safety of the area and that all safety devices and guards have been replaced
9 Return system functions to a preset state.
10 Remove Lock-Out Tag-Out equipment from main disconnect.
11 Locate main power supply (Service) disconnect feeding System Cabinet and re-apply
power.
12 Perform system startup. See Energizing System on page 25.

Table of Contents