Page 268 Chapter 12 - - Set Preferences
Adding A New Detector Identified By Name
Figure 12-13 System - Detector Management- Add By Name
1. Press the [Add] button on the Detector Registration screen.
2. Enter the mac address of the detector you wish to register to the system.
Please detach the detector battery to find the mac address.
3. Chose “Name”for the Select Identification Type.
4. Enter the new Digital Detector Name.
5. Chose to make the new Digital Detector “Associated”or
”Non-Associated”.
If the detector is registered as associated, there will be a message
window for user to confirm that “Detector association status would
change for already present associated detector” if there is already an
associated detector registered to the system.
If the association is successful, the detector will be marked as
associated detector with an asterisk (*) prefixed to the detector name in
the detector registration window. Any previous associated detector in the
system will be automatically changed to non-associated.
If the association failed, the detector information will be stored in the
registered detector list and will not be marked as associated detector.
6. Press [Save]. The new Digital Detector is now registered with the system.
7. Label the new Digital Detector with the Detector Name.