DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Chapter 1 - - Safety and Service Page 29
4.5.1 Personnel Requirements
4.5.2 Overview
This procedure can be performed with:
• Fluke tester
4.5.3 Tools and Test Equipment
Note: The equivalent means that the tool is designed to be capable of performing tests to both the
standard of IEC 62353 and the earth leakage current of IEC 60601-1.
RISK OF ELECTRIC SHOCK.
VOLTAGE PRESENT.
THE SYSTEM MUST BE DISCONNECTED FROM THE FACILITY MAINS INCLUDING THE A1
PANEL GROUND WIRE.
4.5.4 Required Conditions
4.5.5 Set up tools
• How to connect the Tool
1). Connect power cord
2.) Connect red test lead to red input jack (V/Ω/A)
3.) Connect black test lead to black input jack
Personnel Requirements Preliminary Reqs Procedure Finalization
1 Not Applicable 30min 10min
Item Qty. Effectivity Part# Manufacturer
Fluke ESA612 or equivalent 1 - - Fluke
Condition Reference Effectivity
All sub-system covers must be removed to access the
grounding points within each sub-system.
1-