IRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Page 290 Chapter 12 - - Set Preferences
12. When finished adjusting parameters, click [Save] or [Save To Multiple].
[Save] applies the parameters to the currently selected patient size.
A message appears: “Save changes?”
Click [Yes].
[Save To Multiple] applies the parameters to multiple patient sizes within the
selected anatomy.
The Save to Multiple screen appears.
Select the
Patient Size to save the new look to.
Click [Save].
13. Click [Close] on the Image Processing Preference Editor.
14. Log off and reset the system for changes to take effect. Refer to
Chapter 5:
Installation & Configuration
for more information.
Grid Status
If grid status is <Out>, Grid Line Removal algorithm will be not
used. If grid status is <In> and at the same time Grid Line
Reduction selection in Utility is On, Grid line reduction algorithm
will be used.
TE Settings
See Tissue Equalization section for range and effect of TE sliders.
Refer to Tissue Equalization Overview for more information.
Note: The combined total of the TE Under-penetrated and TE Over-penetrated
areas cannot exceed 100%. The slider will automatically stop and a
message appears: “The Total Area cannot exceed 100%. Reduce the
(Under- or Over-) Penetrated area to proceed.”
Parameter Definition