GE HEALTHCARE
D
IRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Page 302 Chapter 12 - - Set Preferences
Enable or Disable Technical Mode
Technical Mode is a special setting that configures the system for image
quality testing and detector calibration. When Technical Mode is enabled, DI is
calculated using the central area of the image (512 pixels x 512 pixels)
regardless of the imaged anatomy.
IMPORTANT!:Do not acquire exam images when Technical Mode is enabled. The use
of Technical Mode is reserved for image testing and calibration purposes only.
The checkbox (Figure 12-39) indicates if Technical Mode is enabled. Check or
uncheck the box as appropriate and press [Save] to apply the change and close
the screen.
Figure 12-39 Techinical mode checkbox