IRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Page 314 Chapter 12 - - Set Preferences
Add or Edit View
Note: The screens and process to edit a view are the same as for adding a view.
Note: This process has multiple screens.
From the Exam Menu:
1. Select the Category and Exam to add or copy the new view to.
2. Click add button [ ] or edit button [ ] of View column.
The Add View (or Edit View) screen appears.
Figure 12-46 Add (or Edit) View
3. Type a
View Name. Do not use spaces. Use underscores ( _ ) to separate
words.
4. From the
Position View After drop-down list, select the placement of the
view within the exam.
5. Select the
Anatomy.
Note: This will determine the image processing for the view. Be sure to select the appropriate
anatomy for the body part imaged.
6. Select the Default Position.
Options are:
Head up - Patient is in normal head first position on table relative to
detector.
Head down - Patient is on the table feet first.