GE HEALTHCARE
DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Chapter 5 - - Installation & Configuration Page 55
6.) Connect the monitor's display cable connector to Dock’s DVI-D port.
7.) Connect the ethernet cable from Dock’s ethernet port to Hospital's network (If required).
Note: Ethernet PoE port shall be only used to connect wireless AP.
8.) Insert the image workstation into dock’s charging slot.
Note: Insert the workstation with the GE side facing the front side of dock to ensure proper
orientation.
9.) Connect monitor's power cable to Hospital main power.
10.) Connect power plug to Hospital main power.
Hospital Network
PoE Port