General procedures
LOGIQ C9 Series – Basic Service Manual 4-9
5482298-100 English Rev. 5
Power off (continued)
• Cancel
Use this button to exit from the System-Exit menu and
return to the previous operation.
• Exit
(Only available when logged in as GE Service with Service
Dongle)
Select this button when you want to exit to the Windows
Desktop.
NOTE: If you need to restart LOGIQ C9 Series when logged on to
the Windows Desktop, ensure that you do a complete power
down (Shut Down). This is required to power up the Front
End Processor.
Figure 4-2. System - Exit menu (when logged in using Service
Dongle)
System shutdown
Disconnect the Mains Power Cable is necessary. For example:
Relocating the scanner.
CAUTION
DO NOT unplug and/or transport the unit until after the power
off sequence has been completed. Failure to do so may result
in corrupted patient files.