Loading the software
LOGIQ C9 Series – Basic Service Manual 8-13
5482298-100 English Rev. 5
Loading the software
Purpose of this section
This section describes how to reinstall software on LOGIQ C9
Series.
Customer provided prerequisite
Formatted and labelled media for Images storage.
Formatted and labelled media for Patient Archive and
Presets (User Defined Settings).
Password for the user ADM.
Data Management - moving all images
Move the images to a remote server or to removable media.
For instructions, please see “Disk management” in the User
Manual/User Guide.
CAUTION
An error, or a power loss may occur.
Always backup the Patient Archive and the Presets (System
Configurations) before loading the software!
In order to complete a successful restore of the Patient
Database, as needed after a hard disk replacement, or if all the
content on the hard disk has been erased, the images must be
moved away from LOGIQ C9 Series before doing backup of
the Patient Database.
Depending on the location set-up, either move the images to a
remote server or to removable media like DVD or CD discs.