EasyManua.ls Logo

GE LOGIQ C9 Series - Electrical Specifications

GE LOGIQ C9 Series
247 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
System Setup
3-20 LOGIQ C9 Series – Basic Service Manual
5482298-100 English Rev. 5
Electrical specifications
Verification of the system’s voltage setting
Verify that the mains voltage specified for the LOGIQ C9 Series
is available on-site.
The voltage setting for the unit is found on a label near the
Mains Power Circuit Breaker on the rear of the system.
Refer to the latest revision User Manual for a fill list of product
labels.
Electrical specifications for LOGIQ C9 Series
In the table below, the electrical specifications for LOGIQ C9
Series includes monitor and on board peripherals.
WARNING
Connecting a LOGIQ C9 Series to the wrong voltage level will
most likely destroy it.
Table 3-7: Electrical specifications for LOGIQ C9 Series
Part
Number
Description Voltage Tolerances
Power
consumntion
Frequency
5489329 LOGIQ C9 Premium for
China
220-240V ±10% Max.950VA 50/60 Hz
5489330 LOGIQ C9 Expert for
China (with 4D)
220-240V ±10% Max.950VA 50/60 Hz
5489331 LOGIQ C9 Expert for
China
220-240V ±10% Max.950VA 50/60 Hz
5489332 LOGIQ C9 Expert for
India
220-240V ±10% Max.950VA 50/60 Hz

Table of Contents