Gathering Trouble Data
LOGIQ C9 Series – Basic Service Manual 7-5
5482298-100 English Rev. 5
Collect a Trouble Image with Logs (continued)
Use a log to if the user wants to send information to GE. There
are two methods to follow:
The subsequent file is compressed and time stamped. The
screen capture is a bitmap which eliminates the possibility of
artifacts from compression.
To save log files and an image snapshot, press Alt-D, select
the save logs function, and fill in the form. Use a USB flash
drive or CD-R to store the log(s). When you fill out the form,
fill in the system behavior you saw, plus any other
comments. Fill in the check box if the system crashed.
To save log files and an image snapshot, press Alt-D.
Describe the issues, then select to store logs to
D:\SERVICE. Select ContactGE in Serivce Menu, then
complete the form of Contact Information, Problem Type
and Problem Description and select Send.