Disassembly/Re-assembly
LOGIQ C9 Series – Basic Service Manual 8-11
5482298-100 English Rev. 5
Cleaning Procedure
1. Pull out the two air filters from both sides of the power box.
Figure 8-4. Pull out Air Filter
2. Dust the filter with a vacumm cleaner and/or wash it with a
mild soapy solution.
If washed, rinse and dry the filter before re-installation.
3. Put back the two air filters to both sides of the power box.