Safety considerations
LOGIQ C9 Series – Basic Service Manual 1-19
5482298-100 English Rev. 5
Probes
Follow these guidelines before connecting a probe to the
Ultrasound system:
Inspect the probe prior to each use for damage or
degradation to the:
• housing
cable strain relief
• lens
• seal
• connector pins
• locking mechanism
Do not use a damaged or defective probe.
Never immerse the probe connector or adapter into any
liquid.
The system has more than one type of probe port. Use the
appropriate probe port designed for the probe you are
connecting.
Peripherals
Refer to the Patient Safety Environment section of the User’s
Manual for peripheral isolation information.