DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
8-84 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-33-3 Removal procedure (cont’d)
5.) Mount ASIR ROM to new SYSCONML Assy. Refer to the Figure 8-119.
6.) Confirm mounting direction and pin number.
NOTE: It is important to swap ASIR ROM(U23). Because it contains the serial Number and option the
system.
Figure 8-118 Separation of ASIR ROM
Figure 8-119 ASIR ROM mounting direction