GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 9-6 - System maintenance 9 - 9
9-6-2-2 Peripheral/Option Checks
If any peripherals or options are not part of the system configuration, the check can be omitted.
Refer to Section 5-8 "Regional and Peripheral Options" on page 5-25 for a list of approved peripherals.
9-6-2-3 Mains Cable Inspection
9-6-2-4 Optional diagnostic checks
To complete the Ultrasound system checks, view the error logs and run desired diagnostics.
9-6-2-5 View the logs
1.) Review the Ultrasound system error log for any problems.
2.) Check the temperature log to see if there are any trends that could cause problems in the future.
Table 9-5 GE approved Peripheral/Hardware Option Functional Checks
Step Item Description
1
B/W Printer
Verify hardcopy output of the B/W video page printer.
Clean heads and covers if necessary.
2
Color Printer
Verify hardcopy output of the Color video page printer.
Clean heads and covers if necessary.
3
Media
Verify hardcopy output of the Deskjet (Bluetooth) printer.
Clean heads and covers if necessary.
4
DVR Verify record/playback capabilities of the DVR.
5
DICOM Verify that DICOM is functioning properly. Send an image to a DICOM device.
6
Footswitch Verify that the footswitch is functioning as programed. Clean as necessary.
7
DVD-Drive
Verify that the DVD-drive reads/writes properly
(export/recall image in Image Management System = Archive)
Table 9-6 Mains Cable Inspection, as appropriate
Step Item Description
1
Unplug Cord Disconnect the mains cable from the wall and system.
2
Inspect Inspect it and its connectors for damage of any kind.
3
Terminals
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that
no strands may cause a short circuit.
4
Inlet Connector Inlet connector retainer is functional.