DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
9 - 12 Section 9-6 - System maintenance
2.) Pull out the air filter with the frame.
Figure 9-3 Remove the air filter
3.) Dust the filter with a vacuum cleaner and/or wash it with a mild soapy solution. If washed, rinse and
dry the filter before re-installation.
4.) Put back the air filter with both hands and the front cover.
Figure 9-4 Put back the air filter
NOTE: For your convenience or if the air filter is too dirty, replacement filters are available. Refer to
Chapter 9 for the air filter replacement part number.