6.Replacethestainlesssteeltubewiththenotchedend(1)onthesideshownin
thefollowingillustration.
7.Reassembletheprinterassemblytothemainassemblyasdescribedin
“ReplacingthePrinterAssembly”onpage83.
8.CongurethePrintheadResistanceasfollows:
a.OntheMainMenu,pressF5(SystemConguration).
b.PressF6(More)>F6(More)>F5(ServiceSetup).
c.Whenprompted,entertheServicePassword.
Ifyoudonotknowthepassword,contactGEHealthcareTechnical
Support.
d.SelecttheDeviceSettingsbutton.
e.InthePrintheadResistanceeld,type1000
f.PressF6(Save).
9.Performtheapplicablecheckoutprocedures.
Referto“FunctionalCheckout”onpage99.