B-4 MAC
TM
600 Revision G
2047426-002
Electromagnetic Compatibility: Guidance and Manufacturer’s Declaration-Electromagnetic Immunity
Guidance and Manufacturer’s Declaration-
Electromagnetic Immunity
Immunity test EN60601 test level Compliance level Electromagnetic environment-guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast Transient/ burst
IEC 61000-4-4
± 2 kV for power supply lines ± 2 kV for power supply lines Mains power should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s) ± 1 kV line(s) to line(s) Mains power should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply input
lines
IEC 61000-4-11
<5% Ut (>95% dip in Ut) for 0.5
cycles
40% Ut (60% dip in Ut) for 5
cycles
70% Ut (30% dip in Ut) for 25
cycles
<5% Ut (>95% dip in Ut) for 5
sec
<5% Ut (>95% dip in Ut) for 0.5
cycles
40% Ut (60% dip in Ut) for 5
cycles
70% Ut (30% dip in Ut) for 25
cycles
<5% Ut (>95% dip in Ut) for 5
sec
Mains power should be that of a typical commercial or
hospital environment. If the user of the MAC i requires
continued operation during power mains interruptions,
it is recommended that the MAC 600 be powered from
an uninterruptable power supply or a battery.
Power frequency(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at levels
characteristics of a typical location in a typical
commercial or hospital environment.